Heater-Cooler System 3T • Operation
5.14 CP_IFU_16-XX-XX_USA_015
Temperature alarm
◗ Display
➜ Cause
◗ Effects
➜ Measures
◗ The symbol of the corresponding
patient circuit lights up, the alarm tone
sounds (on the control panel)
➜ Maximum deviations between the
patient’s temperature and the water
temperature during heating/cooling
have been exceeded
◗ The pumps of the corresponding circuit
stop.
➜ You can keep the water circuit running
(if neccessary) by continuously
pressing the corresponding Circuit
Start/Stop key.
➜ You may adjust the values for the
maximum deviation (“Delta”) on the
H/C menu.
➜ You may wait until the deviation drops
within the normal tolerance range
again. Then, start the pumps again.
◗ The symbol of the corresponding circuit
lights up, the alarm tone sounds
◗ The actual value temperature display
shows 41°C or more
➜ Temporary temperature deviation
during operation
➜ Excess temperature,
Defective temperature control
◗ The pumps of the corresponding circuit
stop.
➜ Wait a for short time until the
temperature stabilizes at its set
temperature, restart the pumps.
➜ You may reoperate the circuit shortly
by opening the overflow valve, allowing
the water of the corresponding circuit
to run out and by refilling it with cold
water. This solution is, however, only
suitable for an interim period and if no
replacement unit is available, since the
temperature control is out of order.
➜ Pass the heater-cooler on to your
service technician.