Heater-Cooler System 3T • Routine maintenance
CP_IFU_16-XX-XX_USA_015 6.17
6.9 Check lists: maintenance intervals
Use the following check lists to ensure that the full scope of the maintenance and inspection
routines are carried out when they are due.
Check list: daily or every time the device has been used
✓
◗ Dispose of used disposables in accordance with environmental regulations (wear protective
gloves).
◗ Empty the overflow bottle.
◗ Automatic chemical-thermal disinfection in a washer-disinfector at intervals of 7 days is
recommended for the overflow bottle.
◗ Clean the heater-cooler in compliance with the instructions given in chap. 6.2 “Cleaning and
disinfection of surfaces”.
◗ Clean and check all accessories (tubings, blankets, etc.) as described in the separate operating
instructions.
◗ Record the operating hours of the heater-cooler to ensure that the “regular preventive
maintenance check” by an authorized service technician is scheduled when it falls due.
◗ Have defective or damaged components checked immediately by an authorized service
technician.
Check list: additional maintenance intervals
✓
In addition to the maintenance routines and checks which are performed daily:
◗ Every 1,000 operating hours or yearly: Notify the authorized service technician for the regular
preventive maintenance check of the heater-cooler.