Heater-Cooler System 3T • Routine maintenance
CP_IFU_16-XX-XX_USA_015 6.17
6.9 Check lists: maintenance intervals
Use the following check lists to ensure that the full scope of the maintenance and inspection
routines are carried out when they are due.
Check list: daily or every time the device has been used
Dispose of used disposables in accordance with environmental regulations (wear protective
gloves).
Empty the overflow bottle.
Automatic chemical-thermal disinfection in a washer-disinfector at intervals of 7 days is
recommended for the overflow bottle.
Clean the heater-cooler in compliance with the instructions given in chap. 6.2 “Cleaning and
disinfection of surfaces”.
Clean and check all accessories (tubings, blankets, etc.) as described in the separate operating
instructions.
Record the operating hours of the heater-cooler to ensure that the “regular preventive
maintenance check” by an authorized service technician is scheduled when it falls due.
Have defective or damaged components checked immediately by an authorized service
technician.
Check list: additional maintenance intervals
In addition to the maintenance routines and checks which are performed daily:
Every 1,000 operating hours or yearly: Notify the authorized service technician for the regular
preventive maintenance check of the heater-cooler.