Heater-Cooler System 3T • Installation
CP_IFU_16-XX-XX_USA_015 4.1
4 Installation
4.1 Preparing the installation
Apart from the components included in the delivery, you still need the following components for
installing your heater-cooler:
Compatible tubing sets for connecting the heating/cooling blanket and the oxygenator.
!
A replacement for the heater-cooler in emergencies, e.g. a total system failure.
4.1.1 General and technical requirements
Only operate the heater-cooler at an ambient temperature ranging between +10°C and +35°C (50°F
through 95°F).
Store the heater-cooler in a cool and dry place ( 0°C/32°F through 40°C /104°F max.).
The heater-cooler complies with the requirements for protection class 1 (IEC 60601-1). It must be
powered by a separately fused and properly grounded power supply.
The heater-cooler must be connected to a potential equalization point for the patient’s and the
user’s safety.
The electrical installations must be in accordance with the requirements and standards
IEC 60364-7-710, or the corresponding equivalent local regulations.