Heater-Cooler System 3T • System description
CP_IFU_16-XX-XX_USA_015 3.1
3 System description
3.1 General description
The Stöckert Heater-Cooler System is an independent (i.e. independent of the water supply)
3-circuit-heating/cooling system. The three separate water circuits have been developed to serve as:
◗ Two circuits for the patient (for the heating/cooling blanket, the oxygenator),
common temperatures ranging from 2°C to 41°C, tank volume 6 l
◗ Interchangeable heating/cooling circuit for cardioplegia,
temperatures ranging from 2°C to 10°C (cooling) and/or from 15°C to 41°C (heating),
volume 2 x 3 l
If required, the circuits for the patient and the circuit for cardioplegia can be switched off separately, in
order to increase the activated functional group’s heating and/or cooling performance.
3.1.1 Components
The common delivery scope contains the standard components stated below. For further components
(“optional components”), please contact SORIN GROUP DEUTSCHLAND GMBH, or your local
SORIN GROUP DEUTSCHLAND GMBH distributor.
Standard components
◗ Heater-cooler base unit
◗ 3 pcs. tubing connectors 1/2", 90° angle, with venting valve
◗ 3 pcs. tubing connectors 1/2", straight
◗ Potential equalization cable
◗ Operating instructions
Optional components
◗ Heating/cooling blanket (dim. 55 x 150 cm)