Heater-Cooler System 3T • Routine maintenance
CP_IFU_16-XX-XX_USA_015 6.15
6.7 Instructions for handling the tubings
◗ Include the tubings normally used with the heater-cooler in the disinfection procedure for the
heater-cooler described in chap. 6.3.1. To this end, connect the tubings during the disinfection
cycle to the heater-cooler.
◗ Tubings that are used with the heater-cooler must be replaced at least once a year.
◗ Only use tubings that are certified for drinking water systems, such as SORIN part number
75-510-218.