Heater-Cooler System 3T • Routine maintenance
CP_IFU_16-XX-XX_USA_015 6.5
6.2 Cleaning and disinfection of surfaces
Apart from the hygienic aspect, it is essential for the operational safety of the heater-cooler that
the system is clean. Perform the following cleaning routine every time the system has been used.
◗ Close the CAN-jack with the matching cover.
Required material
– ready-to-use disinfectant tissues
The following disinfectant tissues have been tested for material compatibility with the heater-
cooler:
◗ Clean all accessible system surfaces.
➜ Use pre-soaked, ready-to-use disinfectant tissues.
➜ Always wipe off spills (blood, etc.) from the system immediately.
!
➜ Never use cleaning agents containing oil or grease.
➜ Never use cleaning agents containing acetone, as these may cause damage to the plastic
components and painted surfaces.
➜ Avoid using cloths which cause static charge (polyester, etc.).
➜ Liquids must not enter the housing. Therefore, do not use sprays.
◗ Disinfect all accessible system surfaces.
➜ Use pre-soaked, ready-to-use disinfectant tissues.
➜ The disinfectant tissues must be approved by the manufacturer for surface disinfection on
medical devices.
➜ No disinfectant residue must remain on the treated surfaces.
➜ When disinfecting, ensure that no liquids enter the housing.
Product name Manufacturer
Bacillol Bode
CaviWipe Metrex
Super Sani Cloth PDI
Meliseptol HBV B. Braun
Mikrozid AF Schülke & Mayr