Heater-Cooler System 3T • Routine maintenance
CP_IFU_16-XX-XX_USA_015 6.5
6.2 Cleaning and disinfection of surfaces
Apart from the hygienic aspect, it is essential for the operational safety of the heater-cooler that
the system is clean. Perform the following cleaning routine every time the system has been used.
Close the CAN-jack with the matching cover.
Required material
ready-to-use disinfectant tissues
The following disinfectant tissues have been tested for material compatibility with the heater-
cooler:
Clean all accessible system surfaces.
Use pre-soaked, ready-to-use disinfectant tissues.
Always wipe off spills (blood, etc.) from the system immediately.
!
Never use cleaning agents containing oil or grease.
Never use cleaning agents containing acetone, as these may cause damage to the plastic
components and painted surfaces.
Avoid using cloths which cause static charge (polyester, etc.).
Liquids must not enter the housing. Therefore, do not use sprays.
Disinfect all accessible system surfaces.
Use pre-soaked, ready-to-use disinfectant tissues.
The disinfectant tissues must be approved by the manufacturer for surface disinfection on
medical devices.
No disinfectant residue must remain on the treated surfaces.
When disinfecting, ensure that no liquids enter the housing.
Product name Manufacturer
Bacillol Bode
CaviWipe Metrex
Super Sani Cloth PDI
Meliseptol HBV B. Braun
Mikrozid AF Schülke & Mayr