Heater-Cooler System 3T • Operation
CP_IFU_16-XX-XX_USA_015 5.5
Mixing the tank content
Press the key Circuit Start/Stop 31 on the heater-cooler to start the
cooling circuit (cooling tank = blue dot).
The green key LEDs flash alternately.
After 5 minutes, press the key Circuit Start/Stop 31 again to stop the
circuit.
The mixing process is completed.
Disconnect the inlet of the cardioplegia circuit 10 from the inlet of the
patient circuit 8 and connect the required tubings of the water
circuits.
°C °C °C
°C °C °C
1 2
37.6
20.1 20
37
10
5
C