Heater-Cooler System 3T • Appendix
7.10 CP_IFU_16-XX-XX_USA_015
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Stöckert Heater-Cooler System 3T is used exceeds the applicable RF compliance level
above, the Stöckert Heater-Cooler System 3T should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the Stöckert Heater-Cooler System 3T.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.