ElectrolyteA3
Chemicals All chemicals are chemically pure, preferably analytical grade. Per cent
is, where no other is stated, weight per cent.
Butoxyethanol Ethylene glycol monobutyl ether,
CH
3
-(CH
2
)
2
-CH
2
-O-CH
2
-CH
2
OH
Methanol 100% vol., CH
3
OH
Perchloric acid 60%, HClO
4
Health and Safety
Before mixing, read the MSDS thoroughly for the specific components.
The user must follow the instructions for proper work procedure according to the instruction
manual supplied with the equipment.
Note
The product must be disposed of according to local regulations for dangerous
goods.
ElectrolyteD2
1. Mix the phosphoric acid in the distilled water
2. Add ethanol, propanol and urea.
Formula D2
500 ml distilled water
250 ml phosphoric acid
250 ml ethanol
50 ml propanol
5 g urea
Chemicals All chemicals are chemically pure, preferably analytical grade. Per cent is,
where no other is stated, weight per cent.
Ethanol 96% vol., CH
3
-CH
2
OH
Phosphoric acid Ortho phosphoric acid 85%,
(HO)
3
PO
Propanol 2-propanol 100%, CH
3
-CH
2
-CH
2
OH
Urea CO(NH
2
)
2
Distilled water H
2
O