CClleeaanniinngg tthhee cchhaarrggeerr
WWAARRNNIINNGG
• Always wear insulated rubber gloves, in addition to personal protective equipment, when cleaning the SSMMRRTT Pak to
reduce the risk of injury.
• Always disconnect the SSMMRRTT charger from the wall outlet before cleaning to avoid the risk of electrical hazards.
• Do not spray fluid directly onto the SSMMRRTT charger.
• Do not power wash the SSMMRRTT charger.
• Do not use solvents, lubricants, or other chemicals to clean the SSMMRRTT charger unless otherwise directed.
• Do not immerse the SSMMRRTT charger in water or allow water to collect on top of the SSMMRRTT charger to avoid the risk of
electric shock.
To clean the SSMMRRTT charger:
1. Disconnect the SSMMRRTT charger from the wall outlet to avoid electrical hazards during cleaning.
2. Wipe surfaces of the SSMMRRTT charger with a soft cloth dampened with a non-abrasive, hospital disinfectant.
3. Wipe with a cloth moistened with clean water to remove any cleaning chemicals or residue.
4. Dry thoroughly before placing the SSMMRRTT charger back into service.
CClleeaanniinngg tthhee bbaatttteerryy
The SSMMRRTT Pak is designed to be power washable. The preferred method for power washing the SSMMRRTT Pak is with the
standard hospital surgical cart washer or hand held wand unit.
WWAARRNNIINNGG
• Always use any appropriate personal protective equipment while power washing to avoid inhaling contagion. Power
washing equipment may aerate contamination.
• Always wear insulated rubber gloves, in addition to personal protective equipment, when cleaning the SSMMRRTT Pak to
reduce the risk of injury.
• Always use only non-conductive materials to wipe the SSMMRRTT Pak.
• Always avoid excessive water exposure to the SSMMRRTT Pak terminals.
• Always refer to the disinfectant’s Material Safety Data Sheet (MSDS) to verify the pH range. Disinfectants with pH levels
higher than 10.5 may cause the SSMMRRTT Pak housing material to crack.
• Do not directly handle or make contact with the SSMMRRTT Pak terminals while cleaning to avoid the risk of injury.
• Do not immerse the SSMMRRTT Pak in liquid to reduce the risk of electric shock.
• Do not use solvents, lubricants, or other chemicals to clean the SSMMRRTT Pak unless otherwise directed.
CCAAUUTTIIOONN
• Do not steam clean or ultrasonically clean the SSMMRRTT Pak.
• Do not exceed 180 °F/82 °C as the maximum water temperature.
• Do not exceed 240 °F/115 °C as the maximum air dry temperature (cart washers).
• Do not exceed 1500 psi (130.5 bar) as the maximum water pressure. If you use a hand held wand to wash the product,
keep the pressure nozzle at a minimum of 24 in. (61 cm) from the product.
To clean the SSMMRRTT Pak:
1. Remove the SSMMRRTT Pak from the PPoowweerr--PPRROO cot or SSMMRRTT charger.
2. Inspect the SSMMRRTT Pak housing and terminal area for any cracks or damage.
3. Clean the SSMMRRTT Pak with a hospital grade disinfectant with a pH range of 6.5 to 10.5.
4. Rinse the SSMMRRTT Pak thoroughly with clean water to remove any cleaning chemical or residue. Position the SSMMRRTT Pak
to avoid water from pooling near the terminals.
5. Dry the SSMMRRTT Pak thoroughly before insert the SSMMRRTT Pak into a PPoowweerr--PPRROO cot or SSMMRRTT charger.
EN 50 6516-209-001 Rev D.0