Cleaning and Maintenance 43
Annual Maintenance Checklist
PASS
FAIL
Model REF 5920-011-000
Serial Number ___________________
Software Version ___________________
Date Tested ___________________
Tester ___________________
VISUAL
INSPECTION
visual inspection for loose components, alignment of
case and rubber handle
BUTTON AND LED
CHECKS
all buttons functional
all LCDs functional
AC and BATTERY
CHARGE
INDICATORS
AC indicator light ON when AC power present
battery charge indicator shows battery charge status
AC POWER ERROR
no AC error message when AC power removed
AC POWER FAIL AT
START-UP
Unit will alarm at start-up with no AC power (LCD
message and Alarm Indicator Mute button red, no
audible tones). Alarm may be cancelled (LCD
message no AC remains).
PRESSURE
CALIBRATION
VERIFICATION
100 mmHg (± 4 mmHg)
Cuff 1
displayed ___________
measured __________
Cuff 2
displayed ___________
measured __________
250 mmHg (± 5 mmHg)
Cuff 1
displayed ___________
measured __________
Cuff 2
displayed ___________
measured __________