Cleaning and Maintenance 44
475 mmHg (± 9 mmHg)
Cuff 1
displayed ___________
measured __________
Cuff 2
displayed ___________
measured __________
300 mmHg LEAK
TEST
Infrequent (if any) valve and pump activity during 5
minute leak test
TIMER TIMEOUT
Timer times out after time limit occurs indicated by:
audible alarm
Alarm Indicator Mute button flashes red
blinking time indicator
TIMER ALARM
CANCEL
Timer alarm cancel indicated by:
audible alarm stops
Alarm Indicator Mute button flashes red
blinking time indicator
TIMER RESET
Increasing the timer value will reset the timer alarm to the
normal state. Timer alarm reset indicated by:
audible alarm stops
Alarm Indicator Mute button is solid green
time indicator stops blinking
DEFLATE
OPERATION
INTERRUPTION
cuff re-inflates to prior setting (350 mmHg) if
deflation interrupted and deflates fully if deflation
not interrupted
LOW PRESSURE
ALARM
alarm is generated if fill line is disconnected to
simulate a pneumatic leak
TIME, DATE and
SERIAL NUMBER
SET CORRECTLY
time of day correct
date correct
serial number matches serial number on rear
panel label