Service Code Summary 46
6 Service Code Summary
WARNING: DO NOT service this equipment. If you require service,
contact your Stryker sales representative or call Stryker customer service
at 1-800-253-3210. Failure to comply may result in user/patient injury.
Service Alarms Code Description
CALIBRATION ERROR 30 Calibration test detects error.
REAL-TIME CLOCK STOPPED 31 Real-time clock has failed or stopped.
SERIAL INTERFACE ERROR 32 Serial interface fault.
FP NONVOLATILE MEMORY ERROR 33 The nonvolatile memory for the FP processor has
failed a memory test.
FP FILES CORRUPT 34 The nonvolatile memory storing the FP code has
failed a memory test.
A NONVOLATILE MEMORY ERROR 35 The nonvolatile memory for the A code has failed a
memory test.
B NONVOLATILE MEMORY ERROR 36 The nonvolatile memory for the B code has failed a
memory test.
HARDWARE FAULT 37 An internal hardware fault was detected.
POWER SUPPLY FAULT 38 Internal power supply circuit fault was detected.
BLADDER 1 LEAK 40 Bladder 1 leaks or is not operating correctly.
BLADDER 2 LEAK 41 Bladder 2 leaks or is not operating correctly.
VALVE OR AIR PUMP FAULT 42 Fault detected with the operation of the air valves or
the air pump.