21
S
Preparation
for Cleaning
1. Properly prepare an enzymatic detergent according to the detergent
manufacturers recommendations.
2. Dip a clean cloth into the detergent and wipe the entire sterilizable handle
cover.
3. Immerse the sterilizable handle cover in the detergent.
4. Soak the sterilizable handle cover in the detergent for at least 15 minutes.
Cleaning:
Manual Method
1. Remove debris from all surfaces using a soft brush dipped in the
enzymatic solution.
2. Rinse the sterilizable handle cover thoroughly under running lukewarm
water; ush water through all passages.
3. Dry the sterilizable handle cover with a lint free cloth.
Cleaning:
Automated Method
1. Rinse the sterilizable handle cover with treated water at ambient
temperature until there is no visible detergent residue. Continue to
rinse the sterilizable handle cover for a minimum of 30 seconds after all
detergent residue has been removed.
2. Place the sterilizable handle cover in the washer on an incline to facilitate
drainage.
3. Program the washer using the following parameters:
Phase
Recirculation
Time
Water
Temperature
Detergent
Type and
Concentration
(if applicable)
Pre Wash 2 minutes Cold N/A
Enzyme
Wash
2 minutes Hot
Enzymatic
Detergent
Wash 1 2 minutes
Set Point
(66° C)
Regular Detergent
Rinse 1 2 minutes Hot N/A
Dry Phase 7 minutes 115° C N/A
* If necessary, use pressurized air to aid in drying. Visually inspect each
sterilizable handle cover for cleanliness.
Disinfection Not applicable
Drying If necessary, use pressurized air to aid in drying.
Maintenance,
Inspection, and
Testing
Inspect sterilizable handle cover before and after use. Discard the sterilizable
handle cover if it is damaged. Damage includes cracks, chips, holes, tears, or
anything else that might interfere with the integrity of the sterilizable handle cover.
Packaging When lling the autoclave, ensure that the open sides of the sterilizable
handle covers are facing down for pre-vacuum sterilization cycles and are
placed on their side for gravity sterilization cycles (AAMI ST79). Make certain
the sterilizable handle covers are not touching any other items while being
sterilized.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?