the Visum® Blade LED Surgical Light System should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the
Visum® Blade LED Surgical Light System.
b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications
equipment and the Visum® Blade LED Surgical Light System
The Visum® Blade LED Surgical Light System is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the Visum® Blade LED Surgical
Light System can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the Visum® Blade LED Surgical
Light System as recommended below, according to the maximum output power of the communications
equipment.
Rated Maximum
Output Power of
Transmitter
W
Separation Distance According to Frequency of Transmitter
m
150 kHz to 80 MHz
d = 1,2√P
80 MHz to 800 MHz
d = 1,2√P
800 MHz to 2,5 GHz
d = 2,3√P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note:
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is aected
by absorption and reection from structures, objects and people.
10.9 Latex Policy
Stryker Communications guarantees that all Visum LED equipment including accessories, packaging
material, labels, pouches, or other similar items do not contain latex.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Visum and is the answer not in the manual?

Stryker Visum Specifications

General IconGeneral
CategoryMedical Equipment
TypeMobile C-Arm
ManufacturerStryker
ModelVisum
Power Requirements50/60 Hz
ApplicationsOrthopedic, Pain Management

Related product manuals