EasyManuals Logo

Stryker Visum User Manual

Stryker Visum
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
24
S
Sterilization** Individually wrap in two layers of polypropylene wrap, (Kimguard KC600) or
similar.
Place in sterilizer with the following parameters:
Sterilization
Type
Preconditioning
Pulses
Minimum
Temperature
Full
Cycle
Time
Minimum
Dry Time
Gravity-
displacement
Steam Sterilization
– US (wrapped)
N/A 132°C 15
minutes
45
minutes
Gravity-
displacement
steam sterilization
– EU (wrapped)
N/A 134°C 10
minutes
45
minutes
Pre-vacuum steam
sterilization --US
(wrapped)
3 132°C 4
minutes
20
minutes
Pre-vacuum steam
sterilization --EU
(wrapped)
3 134°C 3
minutes
20
minutes
Storage Keep in instrument tray until use.
** Sterilizable camera cover must be sterilized according to ISO17665-1.
The validated gravity dry times exceed the AAMI ST79 standard dry time range for wrapped devices by
15 minutes.
Note: Any modications to these parameters should be based on standard hospital
cleaning/sterilization protocol, and are performed at the discretion of the user, who assumes
the risk for modifying these parameters.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Visum and is the answer not in the manual?

Stryker Visum Specifications

General IconGeneral
CategoryMedical Equipment
TypeMobile C-Arm
ManufacturerStryker
ModelVisum
Power Requirements50/60 Hz
ApplicationsOrthopedic, Pain Management

Related product manuals