65
Final Testing and Inspection
Tools Needed
•T
orque
Wrench
•Fiber Optic Light Cable
•Arthroscopes
•Sidne Voice Activation System
•6
. Serial USB Cable
•BioTek Safety Analyzer
•240 V Transformer
•120V Transformer
•X8000 Fan Direction Fixture
Cosmetic Inspection
1.
Replace the
front
panel,
chassis,
and chas
sis co
ver t
o
fi
t snugly with no
visible gaps.
2.
Check that
the logos
and printin
g on
the back
panel are
clear.
3.
Check that
the logos
and printin
g on
the front
panel are
clear.
4.
All co
nnectors sho
uld be
properly
oriented
with
no visib
le ga
ps.
5.
All surfaces
should be
smooth
and free
of
scratches o
r dent
s.
Bulb Inspection
1.
e bulb should be clean with no
fi
ngerprints, cracks, or thermal paste
compound on the glass surface.
2.
V
erify that
the serial num
ber is
above the
handle on
the bulb
.
3.
Tighten
the female
connectors wi
th the
torque
wrench set
to 18
in-lbs.
4.
e bulb module should slide and lock
into place on the rail plate.
Power on Test
1.
Connect the
X8000
to
the 12
0
V po
wer
supply
.
2.
Press the
power b
utton
on and
verify that
the bulb
does illumina
te.
3.
Insert
the
optical
fi
ber light cable in the
jaw
.
e bulb should ignite and
stay lit.