25
User Manual S200 Open
Update April 2018
For example, in the case of atmospheric pressure configuration, the screen displays the following:
↕→
Air pressure comp.
Switched off
Air pressure
1013↕hPa
The second line represents compensation activation/deactivation: OFF / ON.
By accessing the fourth line, the operator can adjust the pressure value (icon : using the Up or Down
arrows).
7.1.5 Calibration methods
The S200 Open offers two calibration methods, as a saturation %, for the oxygen probe:
Calibration of the final value (calibration at one point 100%)
This method allows the probe to be calibrated quickly. It is a standard method.
Two-point calibration (0 and 100%)
This two-point method calibrates the zero point and the probe slope. This is a more complex
and exact method because a “zero-point solution” is also required.
Select the different calibration methods using the menu: CALIBRATION
↑→
Calibr. oxygen
↑→
2-point Calibr.
Reading 100.1 %
Calibr. 0.0 %
Calibr. 100.0 %
Use the right arrow to select the method.
↕→
1-point calibr.
Reading 100.1 %
Calibr. 100.0 %