43
User Manual S200 Open
Update April 2018
↕→
Analog. Output 1
Analog. Output 2
8.10.1 Assignment of measured values
Advancing through the menus:
BASIC ANALOG / ANALOG OUTPUT / ANALOG OUTPUT1
If analog output 1 is used to transfer the temperature measurement, the following screen will be
obtained:
↕→
Temp. measurement
Range 4-20mA
0/4mA = 0.00°C
20 mA = 14.00°C
Various parameters can be assigned to analog outputs, such as temperature measurement
and an accessible parameter with a digital sensor.
When the analog output is used as a controller, the "No Measurement” setting must be
selected here.
8.10.2 Definition of current active range
The RANGE line allows you to switch between two current range choices:
Conventional 4-20 mA
Extended 0-20mA.
Temp. measurement
Range 0-20mA
0/4mA = 0.00°C
20 mA = 14.00°C
8.10.3 Setting the range
Temp. measurement
Range 0-20mA
↕→
0/4mA = 10.00°C
20 mA = 30.00°C