14 A-96.250.341 / 061015
AMI Powercon
Product Description
2.2. Instrument Specification
Power Supply Voltage: 100–240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
Power consumption: max. 30 VA
Sample
requirements
Flow rate: 5–20 l/h
Temperature: up to 50 °C
Inlet pressure: up to 2 bar
Outlet pressure: pressure free
On-site
requirements
The analyzer site must permit connections to:
Sample inlet: Swagelock tube 1/4” adapter
Sample outlet: FEP flexible tube 6 mm
Max. Altitude: 2000 m above sea level
Measuring
Range
Measuring range Resolution
0.055 to 0.999 s/cm 0.001 s/cm
1.00 to 9.99 s/cm 0.01 s/cm
10.0 to 99.9 s/cm 0.1 s/cm
100 to 1000 s/cm 1 s/cm
Automatic range switching. Values for the cell constant
0.0415 cm
-1
(Swansensor UP-Con1000).
Accuracy ±1 % of measured value ± 1 digit
Temperature
compensation
none
Coefficient
Neutral salts
High-purity water
Strong acids
Strong bases
Ammonia, Ethanolamine
Morpholine
Electronics
housing
Aluminium with a protection degree of IP 66 / NEMA 4X, suitable
for indoor and outdoor installation.
Ambient temperature: - 10 to + 50 °C
Humidity: 10–90% rel., non condensing
Display: backlit LCD, 75 x 45 mm
Pollution degree: Pollution degree 2
Installation category: Installation category II