74 A-96.250.341 / 061015
AMI Powercon
Program List and Explanations
5 Installation
5.1 Sensors
5.1.1 Flow:
None
Q-Flow
Select “Q-Flow” if the sample flow should be monitored and shown
on the display and when using a SWAN flow cell.
5.1.2 Sensor parameters
5.1.2.1 Cell Constant: Enter the cell constant printed on the sensor label.
5.1.2.2 Temp. Corr: Enter the temperature correction printed on the sensor
label.
5.1.2.3 Cable length: Enter the cable length. Set the cable length to 0.0 m if
the sensors are installed in the flow cell on the AMI monitor.
5.1.2.4 Meas. unit: Select the measuring unit as s/cm or as s/m.
5.1.3 Temp. comp:
5.1.3.1 Comp.: Available compensation models are:
none
Coefficient
Neutral salts
High purity water
Strong acids
Strong bases
Ammonia, Eth.am.
Morpholine
5.1.4 Quality Assurance:
5.1.4.1 Level.: Choose the quality level according to your requirements.
0: Off; Quality Assurance is not active.
1: Trend (details see Quality assurance level, p. 52)
2: Standard (details see Quality assurance level, p. 52)
3: Crucial (details see Quality assurance level, p. 52)
4: User; edit user specific limits in menu 5.1.4.2 - 5.1.4.4