AMI Powercon
Maintenance
A-96.250.341 / 061015 61
6.11. Longer Stop of Operation
1 Stop sample flow.
2 Slightly squeeze the ion exchanger bottle.
Thus no water will spill out of the flow cell when loosening the
bottle.
3 Unscrew and carefully remove the ion exchanger bottle with the
exhausted resin.
4 Close the ion exchanger bottle with the screw cover and store it
in a frost-protected room.
5 Screw on an empty bottle.
6 Shut off power of the instrument.