80 A-96.250.341 / 061015
AMI Powercon
Program List and Explanations
5.3.1.2 Sample Flow: Define at which sample flow an alarm should be is-
sued.
5.3.1.2.1 Flow Alarm: Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will
always be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
Note:Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.2.2 Alarm High: If the measuring values rises above the programmed
value E009 will be issued.
Range: 1050 l/h
5.3.1.2.35 Alarm Low: If the measuring values falls below the programmed
value E010 will be issued.
Range: 09 l/h
5.3.1.3 Sample Temp.
5.3.1.3.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E007, is displayed in the mes-
sage list.
Range: 30200 °C
5.3.1.3.25 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E008 is displayed in the message
list.
Range: -10 to + 20 °C
5.3.1.4 Case Temp. high
Alarm high: Set the alarm high value for temperature of electronics
housing. If the value rises above the programmed value E013 is is-
sued.
Range: 3075 °C
5.3.1.5 Case Temp. low
Alarm low: Set the alarm low value for temperature of electronics
housing. If the value falls below the programmed value E014 is is-
sued.
Range: -10 to +20 °C