5.3.2.32.31 Control Parameters
Range for each Parameter same as 5.2.1.43, p. 78
Dosing is controlled by the position of a motor driven mixing valve.
5.3.2.32.22 Run time: Time needed to open a completely closed valve
Range: 5–300 Sec.
5.3.2.32.32 Neutral zone: Minimal response time in % of the runtime. If the re-
quested dosing output is smaller than the response time, no
change will take place.
Range: 1–20 %
5.3.2.32.4 Control Parameters
Range for each Parameter same as 5.2.1.43, p. 78
The relay will be activated repetitively depending on the pro-
grammed time scheme.
5.3.2.24 Mode: Operating mode (interval, daily, weekly)
5.3.2.340 Interval/Start time/Calendar: Dependent on options operating
mode.
5.3.2.44 Run time: time the relay stays active.
Range: 5–6’000 Sec
5.3.2.54 Delay: during run time plus the delay time the signal and control
outputs are held in the operating mode programmed below.
Range: 0–6’000 Sec
5.3.2.6 Signal Outputs: select the behavior of the signal outputs when the
relay closes. Available values: cont., hold, off
5.3.2.7 Output/Control: select the behavior of the control outputs when the
relay closes. Available values: cont., hold, off
The relay will be switched via the Profibus input. No further param-
eters are needed.
5.3.2.32.1 Actuator = Motor valve
5.3.2.1 Function = Timer:
5.3.2.1 Function = Fieldbus: