AMI Powercon
Default Values
A-96.250.341 / 061015 89
Case Temp. high:..................................................................... 65 °C
Case Temp. low: ........................................................................ 0 °C
Relay 1/2 Function:.......................................................................... limit upper
Parameter:.................................................................... Conductivity
Setpoint: .................................................................................30 mS
Hysteresis:...............................................................................10 µS
Delay: ....................................................................................30 Sec
If Function = Control upw. or dnw:
Parameter:...................................................................... Cond 1(sc)
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: .............................................120/min.
Settings: Control Parameters: Setpoint:..............................30 mS
Settings: Control Parameters: P-band: ................................10 µS
Settings: Control Parameters: P-band: .................................1 mS
Settings: Control Parameters: Reset time:...........................0 Sec
Settings: Control Parameters: Derivative Time: ...................0 Sec
Settings: Control Parameters: Control Timeout:................... 0 Min
Settings: Actuator: ................................................ Time proportional
Cycle time: ............................................................................ 60 s
Response time: ..................................................................... 10 s
Settings: Actuator ......................................................... Motor valve
Run time: ............................................................................... 60 s
Neutral zone: ...........................................................................5%
If Function = Timer:
Mode:.................................................................................... Interval
Interval: ................................................................................ 1 min
Mode: ........................................................................................daily
Start time: ........................................................................ 00.00.00
Mode:..................................................................................... weekly
Calendar; Start time: ....................................................... 00.00.00
Calendar; Monday to Sunday:.................................................. Off
Run time: ...............................................................................10 Sec
Delay: ......................................................................................5 Sec
Signal output:............................................................................. cont
Output/Control: .......................................................................... cont
Input: Active............................................................................ when closed
Signal Outputs ........................................................................... hold
Output/Control .............................................................................. off
Fault.............................................................................................. no
Delay .....................................................................................10 Sec