21
E
Changing the probe
The Hi_Cal enables the use of di󰀨erent probes and accessories. To change
the probe, simply unscrew the lever and insert the new probe.
Important :
Check if the eccentric probe is correctly positioned
Clamp it rmly
Make sure to use probes with weight value compatible with the
weight adjustment selector
The probe constant must be checked after any probe change or modication
of the probe position.
The Hi_Cal cursor has been balanced for a standard 2mm ball probe. For
heavier probes such as depth measuring probes, the cursor balance must be
set in the UP position.
Foot pedal input
A foot pedal may be connected to act as PRINT key.
Other functions may be assigned to this external contact (see «Setting and
parameter key chapter»).