24
Technical specications Hi_Cal 150 / Hi_Cal 300 / Hi_Cal 450
Measuring range 150mm | 300mm | 450mm
Total measuring range 155mm | 320mm | 470mm
Resolution 0.01 | 0.001 | 0.0005mm
0.005 | 0.0005 | 0.0002IN
Accuracy [µm] 2.5 + L [mm] / 175 [mm]
(with standard probe, at 20 +0.5 °C)
Repeatability (+2s) 2 µm | 0.00008 IN
Measuring sensor max speed 3m/s
Motor max speed (full range) 0.1m/s
Measuring force 0.2 - 0.3 N
Autonomy (using / standby) 20 / 50 heures
Weight 4.0 kg | 4.7 kg | 6.8 kg
Weight of standard probe 7.5g, possibility of using 0-18g
probe, (0-25g, with measuring
force adjustment)
Data input / output RS232-C | USB | Bluetooth
®
Number of measurements / second
In normal mode 10 mes./s.
In diameter/min/max mode 33 mes./s.
DC power supply 9V DC, 1660mA
Battery 4xAA NiMH 2000 mAh
Operating temperature +10° to +40°C
Storage temperature -20° to +45°C
Thermal expansion coecient (same value than steel) (11.5+1.5) x10
-6
[1/°C]
Measuring system Sylvac inductive system, patented