EasyManua.ls Logo

Teac TD-240A - Calibration Methods; Equivalent Input Calibration Overview

Teac TD-240A
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.CALIBRATION
20
4. CALIBRATION
"Calibration" refers to an operation whereby matching between the TD-240A and a
strain gauge sensor is obtained. The TD-240A uses the two calibration methods as
described below.
Equivalent Input Calibration
This approach uses no actual loads but key entry of the rated output value of the strain
gauge sensor (mV/V) and the rating value (value to be displayed).
This method is simple and employed when actual loads cannot be applied.
For example
Gain will be automatically decided by registering the values indicated as follows:
 for load: 2.001mV/V - 100.0kgf
 for pressure: 2.002mV/V - 10.00kgf/cm2, and
 for torque: 2.502mV/V - 15.00kgf•m.
TD-240A
Strain gauge sensor
Rated output value mV/V
Indicated value
A data sheet will be attached to a strain gauge sensor you buy.
On the data seat, the following values are listed.
Capacity ................................... Load (unit: Kg, t, etc.)
Rated Output............................ Voltage (unit: mV/V)
Non-Linearity Hysterisis
Input ResistanceOutput Resistance
Zero Balance, etc.
The Capacity and the Rated Output are necessary values for the
equivalent input calibration. Input these two values to TD-240A.

Table of Contents

Related product manuals