15.SPECIFICATIONS
85
15. SPECIFICATIONS
15-1. Analog Section
Bride Voltage DC 10V ± 10%
DC 2.5V ± 10%
Output current of max 30mA
Changable by setting key Pad
Signal input range - 3.0 to 3.0mV/V
Equivalent input calibration range 0.5 ~ 3.0mV/V
Equivalent input calibration error < 0.1%FS(0.5mV/V input)
Actual load calibration range 0.5 ~ 3.0mV/V
Zero adjustment range 0 ~± 2.0mV/V
Analog input signal sensitivity 1μV/count
Accuracy Non-linearity:< 0.02%FS(of 3mV/V input)
Zero drift :< 0.5μV/ ℃
Gain drift :< 25ppm/ ℃
A/D converter 100 times /sec. Resolution:16 bits (binary)
Analog filter 4Hz,10Hz,100Hz(Initial),3kHz
Changable by setting key pad
Peak hold function (high-speed analog hold system)
Operation response speed:Approx. 1kHz
(Sin wave : 3mV/V input,
Analog Filter.3kHz)
Accuracy: < 0.1%FS
Reset time:< 50μS