2. General
2022-01 HYDROFLEX PLUS No. 30199906 Rev. No. 1.1 19
<= 4 % CV
Measured under following conditions:
8-way manifold, 300 µl, dispensing rate 3 (250 µl/s/well), wash buffer, Greiner
96 well flat bottom plate
Measured under following conditions:
<= 2 µl / well (Greiner 96-well flat bottom plate)
- Crosswise aspiration, 8-way manifold, aspiration rate: 3, aspiration time: 4 s,
head speed: 10 mm/s, wash buffer.
- Residual volume of <= 2 µl / well cannot be guaranteed if a 16-way manifold is
used and if not, all wells are filled with liquid.
<= 5 µl / well (Greiner 96-well round bottom and V-shaped bottom plates)
- Single aspiration, 8-way manifold, aspiration rate: 3, aspiration time: 4 s, head
speed: 10 mm/s, wash buffer.
- Residual volume of <= 5 µl / well cannot be guaranteed if a 16-way manifold is
used and if not, all wells are filled with liquid.
<= 1 ppm
The Carry over between wells <= 1 ppm cannot be guaranteed if not all wells
are filled with liquid.
Auto-sensing
100 - 120 V or 220 - 240 V, 50/60 Hz
Width: 275 mm, Depth: 366 mm, Height: 180.5 mm
(Width: 10.8 inches, Depth: 14.4 inches, Height: 7.1 inches)
15°C to 35°C (59°F to 95°F)
-20°C to 60°C (-4°F to 140°F)
20 % to 80 % non-condensing
10 % to 80 % non-condensing
Electronic waste (infectious waste)