5. Programming Procedure
2022-01 HYDROFLEX PLUS No. 30199906 Rev. No. 1.1 45
5.2.6 Process Step: Soak
Select required soaking time using -/+ keys
(0 60 min; 1 - 59 seconds )
Press Chg. to select minutes or seconds.
Select and confirm soak step (SOAK).
CX = Cycle Number
PX = Program Step Number
CX PX SOAK
- + EXIT OK
TIME: 0 min 10 s
- + Chg. OK
SHAKE OFF
- + EXIT OK
The shaking speed can be selected
(OFF, LOW, MEDIUM or HIGH).
CX PY SOAK
- + EXIT OK
After the process step has been defined the initial
display appears.
CX = Cycle Number
PY = Program Step Number incremented by 1
Press OK to define another soak step
or press or + to select the next program step,
Program End or a new cycle.
5.2.7 Process Step: User Prompt
Select and confirm User Prompt strip on
required position in current program.
CX PX USER PR.
- + EXIT OK

Table of Contents