8. Maintenance and Cleaning
2022-01 HYDROFLEX PLUS No. 30199906 Rev. No. 1.1 59
Rinse Night
This procedure is used to thoroughly rinse the liquid system and then soak the
manifold in the prime tray full of solution at the end of operation.
Use only distilled water with this procedure.
RINSE SOLUTION OK?
< > YES
Confirm that the rinsing solution
and tube are in the bottle
Select and confirm
RINSE NIGHT procedure
Select the required channel
(1 - 4, depending on instrument configuration)
Press OK and the wash system is rinsed
Confirm that the
waste bottle is empty
The prime tray is filled with liquid and the
manifold is lowered into it.
Press Stop to abort the rinse
procedure.
Do not switch the instrument off.
END appears on the display when the
procedure is finished. The manifold
remains immersed in the prime tray until
End is pressed. Press END to finish the
rinse procedure and the firmware returns
to the Program menu.
Press OK to enter the Rinse menu.