TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date:
09/05/06
17
As the device does not recognize difference between effort reduction and apnea,
generically this last term is used. For default the apnea time is established in 15
seconds, but it can be modified at 5, 10, 30 or 60 seconds (Menu key). The backup
ventilation for ADULT and PEDIATRIC category is made by volume or pressure
mode. In NEONATOLOGY it is made with pressure (PCV).
Common Controls to all Modes
f: Regulation of the ventilator frequency. Disabled in PSV, CPAP, MMV and PSV+VT
Assured.
Vtr: Regulation of the inspiratory flow sensitivity.
Ptr: Regulation of the inspiratory pressure sensitivity.
FIO
2
: Regulation of the fraction of inspired oxygen in the gas mixture.
PEEP/CPAP: To regulate a continuous positive pressure in the breathing circuit.
Manual inspiration: To start an inspiration.
Stand by: To suspend the operation of the ventilator without suppressing the set data.
Nebulization: To begin a period of flow to the nebulizer.
Alarm limits and related keys
-High Inspiratory Pressure
-Low Inspiratory Pressure
-VT high/low
-f max: High breathing frequency
-Loss of PEEP: 2, 4, 6 cm H
2
O.
-Apnea Time: 5, 10, 15, 30, 60 seconds.
-Silence: To suppress the alarm sound up to 60 seconds.
-Selection and Enter: Keys to increase or to lower a
parameter and to accept a selected value.
-Reset: To return some action in course.
-Ctrl: To combine functions with other key.