TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Review : 04
Date:
09/05/06
38
Sensor Inspection
Proportional valves:
AirVal:
1)-Press the Reset key.
At that moment, the Calibration Menu is displayed on the screen.
2)-Choose the option “Air Val” with the selection
keys
3)-Put target value in 5 L/m and press the fiO
2
key.
4) - The valve must be open.
5)-Press the Reset
key to finish and press one more time to go back to the calibration menu.
OxyVal:
Idem AirVal