TECHNICAL AND MAINTENANCE MANUAL
Graph VENTILATOR
Final Control Neumovent Graph Ventilator
- Nebulizer: Check operation.
- Manual trigger: Check operation.
- O2 100%: Check operation.
Lower the alarm limit to the peak pressure level read.
Check alarm activation Reset
Return the alarm limit to 40 cm H2O
Rise the minimum pressure alarm limit to the peak pressure level read.
Check alarm activation Reset
Return the alarm limit to 5 cm H2O
With the initial parameters, lower the High Vt limit to 500 ml.
Check alarm activation. Reset.
Return the limit to 750 ml.
Rise the Low Vt alarm limit to 550 ml.
Check alarm activation. Reset.
Return the limit to 250 ml.
Lower alarm limit to 15 per minute.
Check alarm activation. Reset.
Return alarm limit to 30 per minute.
Close the main supply valve.
Open the supply valve again. Reset.
Oxygen
Follow the procedure indicated for air supply.