EasyManua.ls Logo

Tektronix 2215A

Tektronix 2215A
206 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
b. Set the calibration generator to produce a 20-mV
signal.
c. Set the bottom of the signal on the center horizontal
graticule line using the Channel 2 POSITION control.
d. Set the Channel 2 Input Coupling switch to AC.
e. CHECK— Display is centered about the center hori
zontal graticule line.
f. Move the cable from the CH 2 OR Y input connector to
the CH 1 OR X input connector. Set the VERTICAL MODE
switch to CH 1.
g. Repeat parts c through e using the Channel 1 controls.
h. Disconnect the test equipment from the instrument.
9. Adjust Attenuator Compensation (C12, C11, C5,
C4, C62, C61, C55, and C54)
a. Set:
VOLTS/DIV (both) 0.1 V
Input Coupling (both) DC
A SEC/DIV 20
ns
b. Connect the high-amplitude square wave output via a
50-S2 cable and a 50-li terminator to the CH 1 OR X input
connector.
c. Set the generator to produce a 10-kHz, 5-division
display.
NOTE
Use Table 5-4 to identify the correct capacitor fo r
each channel adjustment.
d. ADJUSTThe 10X LF Comp capacitor for best front
corner.
e. Set the CH 1 VOLTS/DIV switch to 1 V.
f. Set the generator to produce a 5-division display.
Adjustment Procedure—2215A Service
Table 5-4
Attenuator Compensation Adjustments
Adjustment Channel 1
Channel 2
10X LF Comp C12 C62
10X Input C C11
C61
100X LF Comp C5
C55
100X Input C
C4 C54
g. ADJUSTThe 100X LF Comp capacitor for best front
corner.
h. Replace the 50-Q cable and 50-fi termination with the
10X probe (supplied with the instrument).
i. Set the CH 1 VOLTS/DIV switch to 10 mV and the A
SEC/DIV switch to 0.2 ms.
j. Insert the tip of the X I0 probe into the PROBE AD
JUST output jack.
k. Adjust the X I0 probe compensation for a flat top on
the waveform. Refer to the instructions supplied with the
probe for details of compensation adjustment.
I. Set the CH 1 VOLTS/DIV switch to 0.1 V or 1 V (1 OX
PROBE).
m. Connect the X I0 probe tip to the high amplitude gen
erator via a probe-tip-to-BNC connector.
n. Set the generator to produce a 1 kHz, 5-division
display.
v
o. ADJUSTThe 10X Input C capacitor for best flat top.
p. Set the CH 1 VOLTS/DIV switch to 1 V or 10 V (XI0
PROBE) and set the generator to produce a 5-division
display.
q. ADJUSTThe 100X Input C Capacitor for the best
flat top.
r. Set the VERTICAL MODE switch to CH 2.
5-9

Table of Contents

Other manuals for Tektronix 2215A

Related product manuals