1-16
30-g, half-sine, 11-ms duration, three shocks per axis in each
direction, for a total of 18 shocks.
Shock (operating and nonoperating)
15 minutes along each of three major axes at a total displacement
of 0.015 inch (0.38 mm) Pop (2.3 g at 55 Hz), with frequency
varied from 10Hz to 55 Hz to 10Hz in one-minute sweeps. Hold
10 minutes at each major resonance, or if no major resonance
present, hold 10 minutes at 55 Hz (75 minutes total test time).
Vibration
-15 °C to + 55°C, ambient temperature measured at the
instrument's air inlet. Fan exhaust temperature should not exceed
+65°C.
Listed characteristics for vibration and shock indicate those
environments in which the rackmounted instrument meets or
exceeds the requlrernents of MIL-T-28800C with respect to Type
III, Class 5, Style D equipment with the rackmounting rear-
support kit installed. Refer to the Standard Instrument Environ-
mental Specification for the remaining performance requirements.
Instruments will be capable of meeting or exceeding the require-
ments of Tektronix Standard 062-2853-00, class 5.
Temperature (operating)
Environmental Requirements
RACKMOUNTEDINSTRUMENT
Meets requirements of Tektronix Standard 062-1860-00.
X-Ray Radiation
Meets Tektronix Standard 062-2862-00. The instrument will not
change control states with discharges of less than 10 kV.
Electrostatic Discharge Susceptibility
Meets MIL-T-28800C; MIL-STD-461B, part 4 (CE-03 and CS-02),
part 5 (CS-06 and RS-02), and part 7 (CS-01, RE-02, and RS-
03-limited to 1 GHz); VDE 0871, Category B; Part 15 of FCC
Rules and Regulations, Subpart J, Class A; and Tektronix Stand-
ard 062-2866-00.
Environmental Requirements (cont)
EMI (electromagnetic interference)
Performance Requirements
Characteristics
Table 1-2 (cont)
Specification-2430 Service