AFG31000 Series Arbitrary Function Generator Service Manual
26
Adjustment procedures
This section describes the procedures necessary to manually adjust the AFG31000 Series Arbitrary
Function Generators. Perform each adjustment when you remove and replace a circuit board.
Purpose
This procedure returns the arbitrary function generator to compliance with its warranted
characteristics as listed in the AFG31000 Series Specifications and Performance Verification
Technical Reference Manual. It can also be used to optimize the performance of the instrument.
After the removal and replacement of an SG board or CPU board due to electrical failure, perform the
adjustment procedure in this section.
NOTE: These adjustments should be performed every 12 months.