Pinpoint Trigge
rs
B Trigger After Delay Time
The A trigger arms the instrument.
Posttrigger a
cquisition starts on the rst B
edge after the trigger delay time.
Triggering
with Reset
You can specify a condition to reset the
trigger system if it occurs before the B trigger
event. Whe
n the reset event occurs, the
trigger system stops waiting for the B event
and returns to waiting for the A event.
Quick Tips
B-trig
ger delay time and horizontal delay time are independent functions. When you establish a trigger condition using
either the A trigger alone or the A and B triggers together, you can also use horizontal delay to delay the acquisition
by an additional amount.