OPERATION
77
OVP/ OCP
The master unit can shut down slave units
when OVP/OCP is tripped on the master unit
(if the slave connector is wired for shut down
on alarm).
OVP/OCP can be independently tripped on
each slave unit, however the shutdown of the
power or output of the unit is disabled. Only
the alarm will be enabled.
Remote monitoring
Voltage monitoring (VMON) and current
monitoring (IMON) are only supported on the
master unit.
The IMON current represents the total current
of the all the parallelized units.
Remote Sense
Please see the remote sense chapter for details,
page 71.
External Voltage and Resistance Control
Voltage/Resistance controlled remote control
can only be used with the master unit.
Slew Rate
Only the master unit’s current or voltage slew
rate settings apply.
The settable current slew rate is double or triple
that of a single unit, depending on the number
of units that are used in parallel.
For example:
The 2260B-30-36 has a maximum settable
current slew rate of 72A/S. When two 2260-30-
36 units are used in parallel, the maximum
settable slew rate setting is 144A/S.