Section 3: Keithley Configuration Utility (KCon) Model 4200A-SCS Setup and Maintenance User's Manual
3-22 4200A-908-01 Rev. A December 2020
Command set comparison
Characteristic
Command set
4145 Emulation
4200A (Normal)
String reported in response to ID query
ID HP4145B 1.1,1.0
KI4200A Vx.x.x (where x.x.x is the
version number)
GPIB data resolution
5 digits
7 digits
Maximum number of sweep points
1024
4096
Possible instrument configurations
8 SMU/VM/VS
Configuration query
Not supported
*OPT? command
Instrument self test
Not supported
SMUs only
Custom A/D control
Not supported
IT4 command options
200 V, 1 A capability
Supported
1.0 pA source/measure-range capability
(with preamplifer on SMU)
Supported
KXCI always starts in the selected mode.
Tools
The options in the Tools dialog box allows you to change Formulator constants, generate technical
support files, and get KCon version and copyright information.
Formulator Constants
The Formulator Constants option allows you to access the default Formulator constants that are
automatically assigned to new Clarius test modules when they are created. You can edit the default
Formulator constants.
Figure 25: Default Formulator constants

Table of Contents

Other manuals for Tektronix Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 4200A-SCS and is the answer not in the manual?

Tektronix Keithley 4200A-SCS Specifications

General IconGeneral
BrandTektronix
ModelKeithley 4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals