EasyManua.ls Logo

Tektronix Keithley 4200A-SCS - Using a Model 590 C-V Analyzer

Tektronix Keithley 4200A-SCS
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 4: Maintenance Model 4200A-SCS Setup and Maintenance User's Manual
4-6 4200A-908-01 Rev. A December 2020
Before initiating a calibration, allow the system to warm up for at least 30 minutes. Clarius prevents
autocalibration if the system is not sufficiently warmed up.
To autocalibrate:
1. Allow the system to warm up for at least 30 minutes.
2. Remove connections to all SMUs in the 4200A-SCS.
3. Open Clarius.
4. Select Tools.
5. Select Instrument Tools.
Figure 27: Clarius Tools dialog box
6. Select SMU Auto Calibration. A warning dialog box is displayed.
7. Select OK. The SMU Auto Calibration dialog box opens, as shown in the following figure.
Figure 28: SMU Auto Calibration dialog box
8. Select Start. A progress bar is displayed.
When autocalibration is complete, the "Auto calibration successfully completed" is displayed.
9. Select Close. Autocalibration is complete.

Table of Contents

Other manuals for Tektronix Keithley 4200A-SCS

Related product manuals