Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2-81
CURSor:VBArs:POSITION<x>
Positions a vertical bar cursor. The unit is specified by the
CURSor:VBArs:UNIts command, and can be in units of seconds or
frequency (Hertz). If the cursor source is an FFT math waveform,
CURSor:VBArs:POSITION is always in Hertz, regardless of the
value set by CURSor:VBArs:UNIts.
NOTE. If Trigger View is active, the query form returns 9.9E37 and
generates event 221 (Settings conflict).
Cursor
CURSor:VBArs:POSITION<x> <NR3>
CURSor:VBArs:POSITION<x>?
<x> specifies which cursor to position. Correct values are 1 and 2.
<NR3> specifies the cursor position in the units specified by the
CURSor:VBArs:UNIts command. The position is relative to the
trigger except when the cursor source is a math FFT waveform.
The cursor position is limited to the graticule whenever an attempt is
made to move it outside the graticule.
CURSor:VBArs:POSITION2 9.00E–6
Positions the second vertical bar cursor at 9 s.
CURSor:VBArs:POSITION1?
Might return 1.00E–6, indicating the first vertical bar cursor is at
1 s.