Command Descriptions
2-96
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
DIAg:RES Ult:FLAg? (Query Only)
Returns the Pass/Fail status from the last diagnostic test sequence
execution (those run automatically at power on, or those requested
through the Service Menu). Use the DIAg:RESUlt:LOG? query to
determine which test(s) has failed.
Calibration and Diagnostic
DIAg:RESUlt:FLAg?
PASS means that the oscilloscope passes all diagnostic tests.
FAIL means that the oscilloscope has failed at least one of the
diagnostic tests.
DIAg:RESUlt:FLAG?
Returns either PASS or FAIL.
DIAg:RESUlt:LOG? (Query Only)
Returns the internal results log from the last diagnostic test sequence
execution (those run automatically at power on, or those requested
through the Service Menu). The list conta ins all modules and module
interfaces that were tested along with the pass or fail status of each.
Calibration and Diagnostic
DIAg:RESUlt:LOG?
Group