Command Descriptions
2- 136
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
HORizontal:DELay:SCAle
Sets the time per division for the oscilloscope window time base.
This command is equivalent to setting SEC/DIV when Window Zone
or Window is selected from the Horizontal menu.
Horizontal
HORizontal:DELay:SCAle <NR3>
HORizontal:DELay:SCAle?
<NR3> is the time per division. The range depends on the oscillo-
scope model. The acceptable values are in a 1–2.5–5 sequence . If the
window time base scale is set slower than the main time base scale,
both the main and window time base scales are set to the window
scale value.
HORizontal:DELay:SCAle 2.0E–6
Sets the window scale to 2 s per division.
HORizontal:DELay:SCAle 9.0E–6
Sets the window scale to 10 s per division. Since 9 s is not a
correct value within the 1–2.5–5 sequence , it is automatically set to
the closest correct value.
HORizontal:DELay:SCAle?
Might return 1.0E–3, indicating that the window scale is 1 ms per
division.
Group
Syntax
Arguments
Examples