Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2- 257
WFMPre:YUNit
For all model and firmware combinat ions except the TDS200 seri es
with a TDS2CMA communications module or a TDS2MM
measurement module, the set form of thi s comm and sets the vertical
units for the re ference wave form specified by DATa:DESTination.
NOTE. It is possible to set a combination of WFMPre:XUNit and
WFMPre:YUNit for a reference waveform that is inconsistent (for
example, seconds with dB or Hertz with volts). The oscilloscope will
not warn you of this condition. The oscilloscope uses WFMPre:XU-
Nit to determine whether the waveform is a YT or an FF T.
For TDS200 oscilloscopes used with a TDS2CMA communications
or a TDS2MM measurement module, the set form is ignored. Refer to
the WFMPre:YMUlt command for equivalent functionality.
The query form returns the vertical units of the waveform specified
by the DATa:SOUrce command, if that waveform is active or
displayed. If that waveform is not ac tive or displayed, the que ry fai ls
and the oscilloscope generates an execution error with event code
2244 (waveform requested is not active).
Waveform
WFMPre:YUNit <qstring>
WFMPre:YUNit?
<qstring> is “Volts”, “U”,or“dB”.
The TDS1000B, TDS2000B, and TPS2000 series have the following
additional arguments:
A for amps
VA for volt-amps
Group
Syntax
Arguments