Command Descriptions
2-60
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
Stops the in-process factory calibration procedure.
CALibrate:CONTINUE (Set Only)
NOTE. You should only use this command in a qualified service
environment. For more inf ormation about the fact ory calibration
sequence, refer to the service manual for your oscilloscope.
Performs the next step in the factory ca libration operation.
Calibration and Diagnostic
CALibrate:CONTINUE
CALibrate:CONTINUE
Performs the next step in the factory ca libration operation.
CALibrate:FACtory (Set Only)
NOTE. You should only use this command in a qualified service
environment. For more inf ormation about the fact ory calibration
sequence, refer to the service manual for your oscilloscope.
Starts the oscilloscope’s internal factory calibration operation. The
calibration operation consists of a sequence of steps. You send the
CALibrate:CONTINUE command to advance to the next calibration
step. The calibration program automatically sets up the oscilloscope
for each calibration step. Use the CALibrate:ABOrt command to
abort the fac tory calibration.
Group
Syntax
Examples