Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2-63
Oscilloscope vertical settings.
CH1?
Might return the following string for channel 1:
CH1:SCALE 1.0E0;POSITION 0.0E0; COUPLING DC;BANDWIDTH
OFF;PROBE 1.0E0
SELect:CH<x>
CH<x>:BANdwidth
Sets or queries the bandwidth setting of the specified oscilloscope
channel. The value of <x> can vary from 1 through 4 for 4-channel
instruments or 1 through 2 for 2-channel instruments.
This command is equivalent to setting the BW Limit option in the
Vertical menu.
Vertical
CH<x>:BANdwidth { ON | OFF }
CH<x>:BANdwidth?
ON sets the channel bandwidth to 20 MHz.
OFF sets the channel bandwidth to the full bandwidth of the
oscilloscope.
In most acquisition modes, full bandwidth is 60 MHz, 100 MHz, or
200 MHz (depending on the oscilloscope model). There are
exceptions.
Returns
Examples
Related Commands
Group