Command Descriptions
2-64
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
TDS1000, TDS2000, TDS1000B, TDS2000B, and TPS2000 Series
At vertical scales from 2.00 to 4.99 mV/div (sensitivity at the BNC;
that is, aft er the probe factor is removed), the full bandwidth is
20 MHz.
TDS200 Series
At vertical scales of 5 mV/div or less (sensitivity at the BNC; that is,
after the probe factor is removed), the full bandwidth is 20 MHz.
When the acquisition mode is Peak Detect, and the vertical scale at
the BNC is 10 mV/div or less, the full bandwidth is also 20 MHz.
CH2:BANDWIDth ON
Sets the bandwidth of channel 2 to 20 MHz.
CH1:BANDWIDth?
Might return OFF. This indicates there is no bandwidth limiting on
channel 1.
CH<x>:COUPling
Sets or queries the input attenuator coupling setting of the specified
oscilloscope channel. The value of <x> can vary from 1 through 4
for 4-channel instruments or 1 through 2 for 2-channel instruments.
This command is equivalent to setting the Coupling option in the
Vertical menu.
Vertical
CH<x>:COUPling { AC | DC | GND }
CH<x>:COUPling?
AC sets the specified oscilloscope channel to AC coupling.
Examples
Group
Syntax
Arguments