Command Descriptions
TDS200, TDS1000/2000, TDS1000B/2000B, TPS2000 Programmer
2-65
DC sets the specified oscilloscope channel to DC coupling.
GND sets the specified oscilloscope channel to ground. Only a flat
ground-level waveform is displayed.
CH1:COUPlING AC
This establishes AC coupling on channel 1.
CH2:COUPlING?
Might return DC. Indicating that channel 2 is set to DC coupling.
CH<x>:CURRENTPRObe
(TDS1000B, TDS2000B, and TPS2000 Series Only)
Sets or queries the sca le factor of the specified channel for current
probes. The value of <x> can vary from 1 through 4 for 4-channel
instruments or 1 through 2 for 2-channel instruments.
You can issue this command when CH<x>:YUNit <QString> is set
to V. Howeve r, this command only affects the readouts when
CH<x>:YUNit <QString> is set to A.
Vertical
CH<x>:CURRENTPRObe { 0.2|1|2|5|10|50|100|
1000 }
CH<x>:CURRENTPRObe?
0.2 sets the specified oscilloscope channel to 0.2X attenuation.
1 sets the specified oscilloscope channel to 1X attenuation.
2 sets the specified oscilloscope channel to 2X attenuation.
5 sets the specified oscilloscope channel to 5X attenuation.
Examples
Group
Syntax
Arguments

Table of Contents

Related product manuals